SignalStar™ Fluorescence Removal Kit #32722
- IHC
Product Information
Product Description
The SignalStar Fluorescence Removal Kit is compatible with SignalStar Oligo-Antibody Pairs, SignalStar Midplex IHC Buffer Kit #29414, and SignalStar Miniplex IHC Buffer Kit #63043 for use in fluorescent multiplex imaging experiments. This product is required to remove fluorescent oligos following the final round of signal amplification and imaging of SignalStar Oligo-Antibody Pairs. This buffer kit includes the reagents required for 10 slides. SignalStar Multiplex IHC Kits & Reagents are not compatible with all of Cell Signaling Technology® products and protocols that are recommended for use in immunohistochemical assays.
Product Usage Information
The SignalStar Fluorescence Removal Kit is intended for use following the SignalStar™ Multiplex IHC Assay Manual protocol or the SignalStar™ Multiplex IHC Assay for Use on BOND RX Fully Automated Research Stainer by Leica Biosystems protocol. Continue to follow all guidance and recommendations from the original protocol when using the SignalStar Fluorescence Removal Kit.
For the SignalStar™ Multiplex IHC Assay Manual protocol:
1. After the second round of image acquisition, soak slides in dH2O for >30 min to gently remove coverslip.
2. Combine reagents to create the SignalStar Fluorescence Removal Solution:
a. For 5 slides: 75 µL 10X dsDNase Buffer, 7.5 µL dsDNase, and 667.5 µL Nuclease-free Water #12931.
b. For 10 slides: 150 µL 10X dsDNase Buffer, 15 µL dsDNase, and 1,335 µL Nuclease-free Water #12931.
3. Cover with parafilm and vortex for 10 sec. Store all SignalStar kit components on ice when preparing solutions. SignalStar solutions should be used promptly once all reagents have been combined for the run.
4. Incubate slides in 150 µL of SignalStar Fluorescence Removal Solution for 2 hr at 37°C. (Use of a covered humidity chamber is highly recommended to prevent evaporation.)
5. Immerse slides in dH2O for 30 sec.
6. Complete any additional staining and imaging steps following your manufacturer’s protocol.
For the SignalStar™ Multiplex IHC Assay for Use on BOND RX Fully Automated Research Stainer by Leica Biosystems protocol:
1. In the BOND RX software by Leica Biosystems, create a copy of the "CST SignalStar Imaging Round 2” protocol.
2. Change the name of the copy to "CST SignalStar Imaging Round 3” with the abbreviated name "CST Rd3.”
3. Select "Show wash steps."
4. Delete Steps 8-63, saving only steps 1-7 for fluorescence removal.
5. Add steps required for any additional staining following your manufacturer’s protocol.
6. Select "CST SignalStar" as the preferred Detection System (Note: Additional staining steps may require further selection of other Detection Systems. Please contact Leica Biosystems for assistance.)
7. Select "Create Protocol.”
8. Click the Save button and then the Yes button to acknowledge the caution message.
9. After the second round of image acquisition, soak slides in dH2O for >30 min to gently remove coverslip.
10. Combine reagents to create the SignalStar Fluorescence Removal Solution:
a. For 5 slides: 185 µL 10X dsDNase Buffer, 18.5 µL dsDNase, and 1,646.5 µL Nuclease-free Water #12931 in 1 BOND Titration Insert.
b. For 10 slides: 335 µL 10X dsDNase Buffer, 33.5 µl dsDNase, and 2,981.5 µL Nuclease-free Water #12931 in 1 BOND Titration Insert.
11. Cover with parafilm and vortex for 10 sec. Store all SignalStar kit components on ice when preparing solutions. SignalStar solutions should be used promptly once all reagents have been combined for the run.
12. In the BOND RX software, create a study and add slides.
13. When "adding slides," use the selections below for Tissue Preparation on BOND:
a. Slide preparation: -- (no value/not required)
b. Dispense Volume: 150 µL
c. HIER: -- (no value/not required)
14. Select "CST SignalStar Imaging Round 3," ensuring that Slide preparation is selected as "--" and HIER is selected as "--."
15. Print labels, add labels to slides, and place slides onto the slide tray.
16. Place 2-3 drops of dH2O onto each slide before adding BOND covertiles.
17. Start the BOND RX run.
Storage
限制使用
除非 CST 的合法授书代表以书面形式书行明确同意,否书以下条款适用于 CST、其关书方或分书商提供的书品。 任何书充本条款或与本条款不同的客书条款和条件,除非书 CST 的合法授书代表以书面形式书独接受, 否书均被拒书,并且无效。
专品专有“专供研究使用”的专专或专似的专专声明, 且未专得美国食品和专品管理局或其他外国或国内专管机专专专任何用途的批准、准专或专可。客专不得将任何专品用于任何专断或治专目的, 或以任何不符合专专声明的方式使用专品。CST 专售或专可的专品提供专作专最专用专的客专,且专用于研专用途。将专品用于专断、专防或治专目的, 或专专售(专独或作专专成)或其他商专目的而专专专品,均需要 CST 的专独专可。客专:(a) 不得专独或与其他材料专合向任何第三方出售、专可、 出借、捐专或以其他方式专专或提供任何专品,或使用专品制造任何商专专品,(b) 不得复制、修改、逆向工程、反专专、 反专专专品或以其他方式专专专专专品的基专专专或技专,或使用专品开专任何与 CST 的专品或服专专争的专品或服专, (c) 不得更改或专除专品上的任何商专、商品名称、徽专、专利或版专声明或专专,(d) 只能根据 CST 的专品专售条款和任何适用文档使用专品, (e) 专遵守客专与专品一起使用的任何第三方专品或服专的任何专可、服专条款或专似专专